Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2134-2009

Stryker Orthopaedics Restoration Gap Plate Screw; Dia. 6.5 mm, Lnth 15 mm; Made in USA, Sterile; Howmedica...

Howmedica Osteonics Corp / initiated 2009-06-26 / Terminated / root cause: manufacturing process

Stryker Howmedica Osteonics Corp. began a recall of Stryker Orthopaedics Restoration Gap Plate Screw; Dia. 6.5 mm, Lnth 15 mm; Made in USA, Sterile; Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430 on . The reason FDA records is "Raw material supplied by a third party used to manufacture various medical devices does not comply with metallurgical requirements outlines in ASTM standard for titanium surgical implants". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2134-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Stryker Orthopaedics Restoration Gap Plate Screw; Dia. 6.5 mm, Lnth 15 mm; Made in USA, Sterile; Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430.
Product code
MEH Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate
Reason for recall
Raw material supplied by a third party used to manufacture various medical devices does not comply with metallurgical requirements outlines in ASTM standard for titanium surgical implants.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Stryker Orthopaedics issued "Urgent Product Recall" letters via Federal Express priority overnight beginning June 26, 2009 to Branches, Hospital Risk Management and Chief of Orthopaedics and Surgeons. All users are requested to examine their inventory, retrieve all affected product and return for reconciliation to Stryker Orthopaedics. Users are also requested to complete the enclosed Product Accountability Form and fax with necessary information to the firm at 1-201-831-6069. For further information, contact Stryker Orthopaedics at 1-201-831-5970.
Quantity in commerce
51,113 total - all sizes.
Distribution
Worldwide Distribution
Product codes and lots
Catalog number: 2080-0015; GAP PLATE SCREWS D36MLE 18-Jul-2013 GAP PLATE SCREWS E2MMLE 21-Jul-2013 GAP PLATE SCREWS D31MLE 21-Jul-2013 GAP PLATE SCREWS D30MLE 21-Jul-2013 GAP PLATE SCREWS E2PMLE 21-Jul-2013 GAP PLATE SCREWS E2NMLE 21-Jul-2013 GAP PLATE SCREWS KM1MLE 23-Jul-2013 GAP PLATE SCREWS KM3MLE 23-Jul-2013 GAP PLATE SCREWS KM0MLE 25-Jul-2013 GAP PLATE SCREWS LH6MLE 25-Jul-2013 GAP PLATE SCREWS LH1MLE 25-Jul-2013 GAP PLATE SCREWS KM2MLE 25-Jul-2013 GAP PLATE SCREWS KM4MLE 25-Jul-2013 GAP PLATE SCREWS LH4MLE 25-Jul-2013 GAP PLATE SCREWS LEYMLE 28-Jul-2013 GAP PLATE SCREWS 7DTMNE 16-Sep-2013 GAP PLATE SCREWS 7DVMNE 16-Sep-2013 GAP PLATE SCREWS MERMH4 04-Dec-2013 GAP PLATE SCREWS MERMH7 04-Dec-2013 GAP PLATE SCREWS MERMH6 05-Dec-2013 GAP PLATE SCREWS MERMH5 08-Dec-2013 GAP PLATE SCREWS METHDN 21-Dec-2013 GAP PLATE SCREWS METHDT 21-Dec-2013 GAP PLATE SCREWS METHDP 28-Dec-2013 GAP PLATE SCREWS METKM8 30-Dec-2013 GAP PLATE SCREWS METKM7 30-Dec-2013 GAP PLATE SCREWS METHDR 06-Jan-2014