Stryker Orthopaedics Restoration Gap Plate Screw; Dia. 6.5 mm, Lnth 50 mm; Made in USA, Sterile; Howmedica...
Stryker Howmedica Osteonics Corp. began a recall of Stryker Orthopaedics Restoration Gap Plate Screw; Dia. 6.5 mm, Lnth 50 mm; Made in USA, Sterile; Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430 on . The reason FDA records is "Raw material supplied by a third party used to manufacture various medical devices does not comply with metallurgical requirements outlines in ASTM standard for titanium surgical implants". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2141-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Stryker Orthopaedics Restoration Gap Plate Screw; Dia. 6.5 mm, Lnth 50 mm; Made in USA, Sterile; Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430
- Product code
- MEH Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate
- Reason for recall
- Raw material supplied by a third party used to manufacture various medical devices does not comply with metallurgical requirements outlines in ASTM standard for titanium surgical implants.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Stryker Orthopaedics issued "Urgent Product Recall" letters via Federal Express priority overnight beginning June 26, 2009 to Branches, Hospital Risk Management and Chief of Orthopaedics and Surgeons. All users are requested to examine their inventory, retrieve all affected product and return for reconciliation to Stryker Orthopaedics. Users are also requested to complete the enclosed Product Accountability Form and fax with necessary information to the firm at 1-201-831-6069. For further information, contact Stryker Orthopaedics at 1-201-831-5970.
- Quantity in commerce
- 51,113 total - all sizes.
- Distribution
- Worldwide Distribution
- Product codes and lots
- Catalog number: 2080-0050; GAP PLATE SCREWS XEHMKE 06-Jul-2013 GAP PLATE SCREWS K1RMLE 22-Jul-2013 GAP PLATE SCREWS K1JMLE 23-Jul-2013 GAP PLATE SCREWS 7DYMNE 16-Sep-2013 GAP PLATE SCREWS 7E0MNE 16-Sep-2013