Z-2144-2009

Corin Unipolar Adapter Sleeve, For use with Corin Unipolar CoCr Heads; Not for use w/ceramic heads; not for...

Howmedica Osteonics Corp / initiated 2009-06-26 / Terminated / root cause: manufacturing process

Stryker Howmedica Osteonics Corp. began a recall of Corin Unipolar Adapter Sleeve, For use with Corin Unipolar CoCr Heads; Not for use w/ceramic heads; not for use w/ Stainless Steel implants. Stryker... on . The reason FDA records is "The raw material used in the production of the Unipolar adaptor sleeve and the Restoration modular products are not affected by the previously mentioned metallurgical non-compliance issue. however, poor process controls...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2144-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Corin Unipolar Adapter Sleeve, For use with Corin Unipolar CoCr Heads; Not for use w/ceramic heads; not for use w/ Stainless Steel implants. Stryker Orthopaedics. Indicated for use in partial hip replacement procedures for patients suffering from pain and disability due to a variety of related bone disease.
Product code
KWL Prosthesis, Hip, Hemi-, Femoral, Metal
Reason for recall
The raw material used in the production of the Unipolar adaptor sleeve and the Restoration modular products are not affected by the previously mentioned metallurgical non-compliance issue. however, poor process controls demonstrated by this vendor has raised concern about their supplied raw material. As a precaution and due to the potential impact on the mechanical properties of the products, St
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Stryker Orthopaedics issued "Urgent Product Recall" letters via Federal Express priority overnight beginning June 26, 2009 to Branches, Hospital Risk Management and Chief of Orthopaedics and Surgeons. All users are requested to examine their inventory, retrieve all affected product and return for reconciliation to Stryker Orthopaedics. Users are also requested to complete the enclosed Product Accountability Form and fax with necessary information to the firm at 1-201-831-6069. For further information, contact Stryker Orthopaedics at 1-201-831-5970.
Quantity in commerce
24
Distribution
Worldwide Distribution
Product codes and lots
Catalog number: 570-000-01: CORIN UNIPOLAR ADAPTOR SLEEVE JW3MHE 25-Apr-2013 CORIN UNIPOLAR ADAPTOR SLEEVE JXNMHE 25-Apr-2013 CORIN UNIPOLAR ADAPTOR SLEEVE 8DHMJE 17-May-2013 CORIN UNIPOLAR ADAPTOR SLEEVE 8DKMJE 19-May-2013 CORIN UNIPOLAR ADAPTOR SLEEVE 8DVMJE 21-May-2013