Restoration Modular hip System, Conical Distal Stem, 14 mm x 155 mm; Straight, Cementless Use Only. Stryker...
Stryker Howmedica Osteonics Corp. began a recall of Restoration Modular hip System, Conical Distal Stem, 14 mm x 155 mm; Straight, Cementless Use Only. Stryker Orthopaedics. Intended to be used for primary or... on . The reason FDA records is "The raw material used in the production of the Unipolar adaptor sleeve and the Restoration modular products are not affected by the previously mentioned metallurgical non-compliance issue. however, poor process controls...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2147-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Restoration Modular hip System, Conical Distal Stem, 14 mm x 155 mm; Straight, Cementless Use Only. Stryker Orthopaedics. Intended to be used for primary or revision total hip arthroplasty, as well as in the presence of sever proximal bone loss.
- Product code
- LZO Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented
- Reason for recall
- The raw material used in the production of the Unipolar adaptor sleeve and the Restoration modular products are not affected by the previously mentioned metallurgical non-compliance issue. however, poor process controls demonstrated by this vendor has raised concern about their supplied raw material. As a precaution and due to the potential impact on the mechanical properties of the products, St
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Stryker Orthopaedics issued "Urgent Product Recall" letters via Federal Express priority overnight beginning June 26, 2009 to Branches, Hospital Risk Management and Chief of Orthopaedics and Surgeons. All users are requested to examine their inventory, retrieve all affected product and return for reconciliation to Stryker Orthopaedics. Users are also requested to complete the enclosed Product Accountability Form and fax with necessary information to the firm at 1-201-831-6069. For further information, contact Stryker Orthopaedics at 1-201-831-5970.
- Quantity in commerce
- 203 - all varieties
- Distribution
- Worldwide Distribution
- Product codes and lots
- Catalog number: 6276-7-014; REST MOD CONICAL DISTAL STEM 14mmX155mm CAXJ826A 27-Jan-2014 REST MOD CONICAL DISTAL STEM 14mmX155mm CAXJ826F 29-Jan-2014 REST MOD CONICAL DISTAL STEM 14mmX155mm CAXJA10E 17-Feb-2014 REST MOD CONICAL DISTAL STEM 14mmX155mm CAXJ826C 17-Feb-2014 REST MOD CONICAL DISTAL STEM 14mmX155mm CAXJA10E 17-Feb-2014 REST MOD CONICAL DISTAL STEM 14mmX155mm CAXJA10E 17-Feb-2014 REST MOD CONICAL DISTAL STEM 14mmX155mm CAXJ826C 17-Feb-2014 REST MOD CONICAL DISTAL STEM 14mmX155mm CAXJ826C 17-Feb-2014 REST MOD CONICAL DISTAL STEM 14mmX155mm CAXJA10A 19-Feb-2014 REST MOD CONICAL DISTAL STEM 14mmX155mm CAXJA10A 19-Feb-2014 REST MOD CONICAL DISTAL STEM 14mmX155mm CAXJ826C 17-Feb-2014 REST MOD CONICAL DISTAL STEM 14mmX155mm CAXJA10C 20-Feb-2014 REST MOD CONICAL DISTAL STEM 14mmX155mm CAXJA10C 20-Feb-2014 REST MOD CONICAL DISTAL STEM 14mmX155mm CAXJA10C 20-Feb-2014 REST MOD CONICAL DISTAL STEM 14mmX155mm CAXJA10D 05-Mar-2014 REST MOD CONICAL DISTAL STEM 14mmX155mm CAXJA10D 05-Mar-2014 REST MOD CONICAL DISTAL STEM 14mmX155mm CAXJA10D 05-Mar-2014