Z-1692-2009

Prolieve Thermodilatation Catheter Kit, a Transurethral Microwave Therapy (TUMT) System, for BPH (Benign...

Boston Scientific Corporation / initiated 2009-07-20 / Terminated / root cause: manufacturing process

Boston Scientific Corporation began a recall of Prolieve Thermodilatation Catheter Kit, a Transurethral Microwave Therapy (TUMT) System, for BPH (Benign Prostatic Hyperplasia) Thermotherapy, Material Numbers... on . The reason FDA records is "Balloon leaks: Boston Scientific has received complaints from the field involving product leaks associated with the anchor balloons and compression balloons. Correspondence between BSC and FDA has led to BSC's decision...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1692-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
Prolieve Thermodilatation Catheter Kit, a Transurethral Microwave Therapy (TUMT) System, for BPH (Benign Prostatic Hyperplasia) Thermotherapy, Material Numbers M0068808022 and M0068808170.
Product code
MEQ System, Hyperthermia, Rf/Microwave (Benign Prostatic Hyperplasia),Thermotherapy
Reason for recall
Balloon leaks: Boston Scientific has received complaints from the field involving product leaks associated with the anchor balloons and compression balloons. Correspondence between BSC and FDA has led to BSC's decision to remove the Prolieve Catheter Kits from the field.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Boston Scientific sent Customer notification letters via Federal Express Priority mail on July 20, 2009. A total of 812 US consignees were notified and a total of 26 consignees in Puerto Rico. The notification instructs the account to check their inventory, remove any affected product from their inventory, and return it to Boston Scientific. The notification also requests the completion of a Reply Verification Tracking Form as a method of documenting the presence or absence of affected product in their inventory. Follow-up communications will be made to the accounts that do not respond to this inquiry.
Quantity in commerce
26,935 kits
Distribution
Nationwide distribution, including Puerto Rico.
Product codes and lots
Material Number M0068808022 with Lot/Batch Numbers: 604951, 604952, 604953, 604954, 604955, 604956, 604957, 604958, 604959, 604960, 605015, 605016, 605017, 605276, 605277, 605278, 605279, 605280, 605281, 605282, 605283, 605284, 605285, 605286, 605287, 605288, 605289, 605290, 605291, 605292, 605293, 605294, 605295, 605296, 605297, 605298, 605299, 605783, 605784, 605785, 605786, 605787, 605788, 605789, 605790, 605791, 606044, 606045, 606046, 606079, 606080, 606081, 606082, 606083, 606084, 606152, 606182, 606183, 606203, 606299, 606300, 606301, 606302, 606303, 606304, 606420, 606421, 606422, 606423, 606424, 606597, 606598, 606599, 606600, 606601, 606602, 606603, 606604, 606605, 606606, 606607, 606781, 606782, 606829, 606830, 606831, 606832, 606833, 606834, 606879, 606880, 607014, 607015, 607016, 607017, 607018, 607019, 607020, 607021, 607022, 607023, 607024, 607025, 607189, 607190, 607191, 607192, 607193, 607194, 607315, 607316, 607317, 607318, 607321, 607322, 607323, 607324, 607325, 607520, 607521, 607522, 607523, 607524, 607525, 607526, 607527, 607528, 607646, 607735, 607736, 607737, 607738, 607739, 607740, 608160, 608161, 608162, 608163, 608164, 608165, 608166, 608167, 608168, 608169, 608575, 608591, 608592, 608593, 608594, 608595, 608596, 608597, 608602, 608680, 608681, 619219, 619221, 619222, 619311, 619312, 619313, 623435, 623436, 627518, 627519, 628502, 628503, 628504, 633348, 633349, 633350, 637211, 637212, 637216, 637217, 637220, 639295, 639296, 642203, 642204, 642549, 642576, 644535, 644536, 644537, 647635, 647636, 650229, 650230, 650232, 650233, 650234, 651870, 651871, 651872, 653449, and M19406. Material Number M0068808170 with Lot/Batch Numbers: 608608, 608609, 608610, 608611, 608612, 651734, 651874, and 651877.