Z-0904-2011
VISU 200 on S8 Floor Stand, Manufactured by Carl Zeiss Surgical GmbH, Germany Medical device: Surgical...
Carl Zeiss Meditec Inc / initiated 2009-05-18 / Terminated / root cause: labeling and UDI
Carl Zeiss Meditec Inc began a recall of VISU 200 on S8 Floor Stand, Manufactured by Carl Zeiss Surgical GmbH, Germany Medical device: Surgical microscope used during ophthalmic surgeries, which... on . The reason FDA records is "Potential Retina Damage-- An error in the user manual may result in misapplied filter and injury during treatment". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0904-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Carl Zeiss Meditec Inc
- Product
- VISU 200 on S8 Floor Stand, Manufactured by Carl Zeiss Surgical GmbH, Germany Medical device: Surgical microscope used during ophthalmic surgeries, which allows real time visualization of the surgical area. Product Quantity Distributed (Int) 2540 units
- Product code
- EPT Microscope, Surgical
- Reason for recall
- Potential Retina Damage-- An error in the user manual may result in misapplied filter and injury during treatment.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling mix-ups
- Action
- Carl Zeiss Meditec Inc sent a SAFETY ALERT letter dated May 18, 2009, via Federal Express to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were reminded to use the correct retinal protection filter (GG475) when using the OPMI VISU 200 system during ophthalmic surgery. Verify the position of the retinal protection filter and inform their staff of the customizable feature. Verify that the correct filter was in place. Customers were asked to complete the acknowledgement form indicating receipt of the alert. Sign and return the form to: Carl Zeiss Meditec in Dublin, California via the enclosed self-addressed, pre-paid Federal Express envelope. For any questions customers should call (925) 557-4422.
- Quantity in commerce
- 2540 units
- Distribution
- Worldwide Distribution including USA (nationwide) and the UK
- Product codes and lots
- User manuals with numbers G-30-1246-e, G-30-1284-e, G-30-1284-en, and G-30-1236-en, affecting all stands.