Stryker Orthopaedics Cancellous Bone Screw; Dia 6.5 mm; Lnth: 55 mm; Sterile: Made in the USA; Stryker...
Stryker Howmedica Osteonics Corp. began a recall of Stryker Orthopaedics Cancellous Bone Screw; Dia 6.5 mm; Lnth: 55 mm; Sterile: Made in the USA; Stryker Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430... on . The reason FDA records is "Raw material supplied by a third party used to manufacture various medical devices does not comply with metallurgical requirements outlines in ASTM standard for titanium surgical implants". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2177-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Stryker Orthopaedics Cancellous Bone Screw; Dia 6.5 mm; Lnth: 55 mm; Sterile: Made in the USA; Stryker Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430. The bone screws are optional fixtures that may be used during knee and hip replacement procedures.
- Product code
- HWC Screw, Fixation, Bone
- Reason for recall
- Raw material supplied by a third party used to manufacture various medical devices does not comply with metallurgical requirements outlines in ASTM standard for titanium surgical implants.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Stryker Orthopaedics issued "Urgent Product Recall" letters via Federal Express priority overnight beginning June 26, 2009 to Branches, Hospital Risk Management and Chief of Orthopaedics and Surgeons. All users are requested to examine their inventory, retrieve all affected product and return for reconciliation to Stryker Orthopaedics. Users are also requested to complete the enclosed Product Accountability Form and fax with necessary information to the firm at 1-201-831-6069. For further information, contact Stryker Orthopaedics at 1-201-831-5970.
- Quantity in commerce
- 51,113 total - all sizes
- Distribution
- Worldwide Distribution
- Product codes and lots
- Catalog number: 2030-6555-1; 6.5 CANCELLOUS BONE SCREW 55mm AMNMLE 18-Jul-2013 6.5 CANCELLOUS BONE SCREW 55mm AMXMLE 18-Jul-2013 6.5 CANCELLOUS BONE SCREW 55mm 3J6MNE 09-Sep-2013 6.5 CANCELLOUS BONE SCREW 55mm MERJA9 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 55mm MET0Y0 14-Dec-2013 6.5 CANCELLOUS BONE SCREW 55mm MET7DP 14-Dec-2013 6.5 CANCELLOUS BONE SCREW 55mm METE5M 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 55mm MHA02E 07-Jan-2014 6.5 CANCELLOUS BONE SCREW 55mm MHA09P 11-Jan-2014
FIRST ARTICLE / WEEKLY