Bio-Rad VARIANT II Hemoglobin A1c Program Recorder Pack with Calibrator/Diluent Set, Model Number 270-2101NU...
Bio-Rad Laboratories Inc began a recall of Bio-Rad VARIANT II Hemoglobin A1c Program Recorder Pack with Calibrator/Diluent Set, Model Number 270-2101NU Manufactured by Bio-Rad Laboratories, Hercules, CA on . The reason FDA records is "Calibration Failure-- due to high intercept values, the device fails calibration and users are unable to run samples". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1013-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Bio-Rad Laboratories Inc.
- Product
- Bio-Rad VARIANT II Hemoglobin A1c Program Recorder Pack with Calibrator/Diluent Set, Model Number 270-2101NU Manufactured by Bio-Rad Laboratories, Hercules, CA
- Product code
- LCP Assay, Glycosylated Hemoglobin
- Reason for recall
- Calibration Failure-- due to high intercept values, the device fails calibration and users are unable to run samples.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- Recall initiated June 15, 2009, with notification initially conducted by phone with Fax Notification follow up domestically. Each subsidiary will be given an urgent notice, subsidiary response form, customer notification, customer acknowledgement form. Each subsidiary is to contact customers and translate notification and response form as needed. Contact the recalling firm for additional information.
- Quantity in commerce
- 613 units
- Distribution
- Product was distributed to 127 customers throughout the US and internationally to Argentina, Australia, Canada, Chile, Czech Republic, France, Germany, Hong Kong, Hungary, India, Italy, New Zealand, Poland, Singapore, Sweden, Taiwan, and the United Kingdom
- Product codes and lots
- Medical device, intended for percent determination of hemoglobin A1c in human whole blood using ion-exchange high-performance liquid chromatography, intended for professional use only.