Clinical Chemistry Total Bilirubin, 6L45-20 and 6L45-40. Used for the quantitative analysis of total...

Abbott Medical / initiated 2009-07-09 / Terminated / root cause: manufacturing process

Abbott Laboratories Inc. began a recall of Clinical Chemistry Total Bilirubin, 6L45-20 and 6L45-40. Used for the quantitative analysis of total bilirubin in human serum or plasma of adults and neonates on . The reason FDA records is "Abbott Laboratories has issued a recall on the Clinical Chemistry Total Bilirubin due to an under-recovery on one clinical neonatal specimen resulting in a value less than the assay linear limit". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1425-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Clinical Chemistry Total Bilirubin, 6L45-20 and 6L45-40. Used for the quantitative analysis of total bilirubin in human serum or plasma of adults and neonates.
Product code
CIG Diazo Colorimetry, Bilirubin
Reason for recall
Abbott Laboratories has issued a recall on the Clinical Chemistry Total Bilirubin due to an under-recovery on one clinical neonatal specimen resulting in a value less than the assay linear limit.
Root cause tag
manufacturing process
FDA root cause
Material/Component Contamination
Action
Abbott Laboratories plans to release a Product Correction letter to customers instructing customers to follow the steps provided in order to revise their Total Bilirubin assay parameters. The letter explained the reason for the recall, along with the patient impact. Customers were to keep the letter as part of their laboratory records and also forward a copy to any other laboratories that they may have forwarded Total Bilirubin kits. U.S. Customers should contact Customer Service at 1-877-4ABBOTT, if they have questions.
Quantity in commerce
11,282
Distribution
Worldwide Distribution -- USA, Guatemala, Chile, Brazil, Canada, Germany, Hong Kong, Thailand, Puerto Rico, Australia, New Zealand, Trinidad & Tobago, Bermuda, Curacao, Cayman Island, and Jamaica.
Product codes and lots
List Numbers: 6L45-20, 6L45-40 Lot Numbers: 64047HW00, 64048HW00, 65084HW00, 65088HW00, 69001HW00, 69002HW00, 72016HW00, 73038HW00, 73057HW00, 75036HW00, 76023HW00, 77014HW00, 69025HW00, 73015HW00, 76038HW00, 77015HW00.