Z-1229-2010
5mm Tibial Augment Block, a component of Ortho Balanced Knee Modular Tibial System. The device is a component...
Ortho Development Corp. / initiated 2009-03-19 / Terminated / root cause: design
Ortho Development Corporation began a recall of 5mm Tibial Augment Block, a component of Ortho Balanced Knee Modular Tibial System. The device is a component of Balanced Knee Modular Tibial System and is... on . The reason FDA records is "A component of knee prosthesis was assembled incorrectly". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1229-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ortho Development Corp.
- Product
- 5mm Tibial Augment Block, a component of Ortho Balanced Knee Modular Tibial System. The device is a component of Balanced Knee Modular Tibial System and is indicated for patients suffering from severe knee pain and disability. The indications for use are: 1) loss of joint configuration and joint function, 2) Osteoarthritis of the knee joint, 3) Rheumatoid arthritis of the knee joint, 4) Post-traumatic arthritis of the knee joint, 5) Moderate valgus, varus, or flexion deformities and 6) Revision procedures where other treatments or devices have failed.
- Product code
- JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Reason for recall
- A component of knee prosthesis was assembled incorrectly.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- E-mail was sent to all agents and distributors on March 19, 2009 with illustrations and instructions on how to detect the upside down assembly and how to correct the problem if detected during surgery. Direct questions to the Ortho Development Corporation by calling 1-801-619-3444.
- Quantity in commerce
- 368 units
- Distribution
- United States (AZ, FL, OR, TX and UT) and Japan.
- Product codes and lots
- Serial Numbers: 567-5101; 567-5102; 567-5103; 567-5104; 567-5105; 567-5106; 567-5201; 567-5202; 567-5203; 567-5204; 567-5205 and 567-5206.