Z-1898-2009

SmartPort CT Port Access System, Catalog number CT80STPD, AngioDynamics Incorporated The SmartPort CT Port...

Boston Scientific Corporation / initiated 2009-07-16 / Terminated / root cause: packaging

Angiodynamics, INC began a recall of SmartPort CT Port Access System, Catalog number CT80STPD, AngioDynamics Incorporated The SmartPort CT Port Access System is indicated for any patient requiring... on . The reason FDA records is "The product may contain an incorrect size catheter and introducer". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1898-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
SmartPort CT Port Access System, Catalog number CT80STPD, AngioDynamics Incorporated The SmartPort CT Port Access System is indicated for any patient requiring repeated access of the vascular system or other selected body site, for delivery of medications, nutritional supplementation, fluids, blood products, and sampling of blood.
Product code
LJT Port & Catheter, Implanted, Subcutaneous, Intravascular
Reason for recall
The product may contain an incorrect size catheter and introducer.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
Notification of the AngioDynamics, Inc. sales force began on July 16, 2009 and they were asked to assist in the return of all affected devices to the Manchester, GA division of AngioDynamics, Inc. Urgent Recall letters were sent to all customers on July 16, 2009, requesting the completion of a FAX BACK form. Those currently in possession of any remaining devices will be provided a Return Goods Authorization (RGA) number and pre-paid shipping boxes for return of the product to AngioDynamics, Inc. All returned devices will be physically and electronically quarantined until final disposition of the devices. All devices will be inspected for damage to the devices and any damaged devices destroyed (scrapped) via the AngioDynamics procedure for Disposition of Nonconformances, procedure. Non-damaged product may be reworked to remove the incorrect catheter and introducer and replaced with the correct catheter and introducer, repackaged and re-sterilized in accordance with the AngioDynamics procedure for Rework. If necessary, follow-up letters, phone calls or visits by the AngioDynamics sales representatives may be used to complete the effective checks. Level A (100%) of consignees to be contacted). Please contact your local representative or Recall Coordinator at 1-706-846-3126, for assistance. You may e-mail your questions to [email protected].
Quantity in commerce
150 kits
Distribution
AL, CA, CO, FL, GA, ID, IL, IN, KS, LA, MI, MO, MN, MS, NC, NY, OH, OR, PA, TN and TX.
Product codes and lots
Lot number: 983533