BARD Simpulse" SOLO System with Soft Splash Shield Tip

Davol Inc., Sub. C.R. Bard, Inc. / initiated 2009-07-30 / Terminated / root cause: manufacturing process

Davol, Inc., Sub. C. R. Bard, Inc. began a recall of BARD Simpulse" SOLO System with Soft Splash Shield Tip Product Code: 0067590. The Simpulse SOLO Suction/Irrigator is designed to provide controlled powered... on . The reason FDA records is "Product sterility may be compromised due to lack of package integrity". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1948-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Davol Inc., Sub. C.R. Bard, Inc.
Product
BARD Simpulse" SOLO System with Soft Splash Shield Tip Product Code: 0067590. The Simpulse SOLO Suction/Irrigator is designed to provide controlled powered irrigation during open orthopedic and general surgical procedures and for the cleansing of burns or wounds. Appropriate fluids for use include fluids with specific gravity less than 1. The pulsatile action of the pump helps to remove blood, tissue debris and foreign matter from the operative or wound site. When connected to a suction source, the device can be used to aspirate fluids from the operative or wound site.
Product code
FHQ Holder, Needle, Gastroenterologic
Reason for recall
Product sterility may be compromised due to lack of package integrity.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Davol issued a Recall Letter dated July 30, 2009 requesting user facilities examine inventory and immediately discontinue use of any product from the affected lot and return it to Davol. Direct questions about the recall to Davol, Inc., Sub. C. R. Bard, Inc. by calling 1-401-463-7000.
Quantity in commerce
1,140 units
Distribution
Nationwide.
Product codes and lots
Lot Number JUSHF083.