Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2026-2009

Icelock Expulsion Valve 551 Model L-551000 The valve has a push button for suction release to release a lower...

Ossur Engineering, Inc / initiated 2009-06-23 / Terminated

Ossur Engineering, Inc began a recall of Icelock Expulsion Valve 551 Model L-551000 The valve has a push button for suction release to release a lower limb prosthesis on . The reason FDA records is "The plastic caps have come off or become loose when the user stepped into their prosthesis". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2026-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ossur Engineering, Inc
Product
Icelock Expulsion Valve 551 Model L-551000 The valve has a push button for suction release to release a lower limb prosthesis.
Product code
ISP Valve, Prosthesis
Reason for recall
The plastic caps have come off or become loose when the user stepped into their prosthesis.
FDA root cause
Process change control
Action
Customers were sent a Safety Notice Voluntary Product Recall, dated June 23, 2009, that advised them to return any unused valves to Ossur for a free replacement. Those using valves in combination with a cushion liner and suspension sleeve should replace immediately if problems are experienced, otherwise replace at next routine service. Customers using valve in combination with an Iceross Seal-In liner should replace ASAP and return used valves to Ossur for free replacement. Customers with questions can call Darlyne Roa 1-800-233-6263 ext 1023743.
Quantity in commerce
4,647
Distribution
Worldwide Distribution
Product codes and lots
Lot CO081010