Applied Medical 15x100mm Kii Optical Access System, Non-Threaded 6/Box, Model Number: C0R36, Labeled Sterile...
Applied Medical Resources Corp began a recall of Applied Medical 15x100mm Kii Optical Access System, Non-Threaded 6/Box, Model Number: C0R36, Labeled Sterile. Disposable single-use devices indicated for use... on . The reason FDA records is "Applied Medical is conducting a coluntary recall of two lot numbers, 1084426 and 1085506, of the 15mm Kii non-threaded optical access system, model C0R36, due to a potential weakness in the cannula tip, possibly...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2226-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Applied Medical Resources Corp
- Product
- Applied Medical 15x100mm Kii Optical Access System, Non-Threaded 6/Box, Model Number: C0R36, Labeled Sterile. Disposable single-use devices indicated for use in general, abdominal, gynecological and thoracic minimally invasive surgeries. Endoscopic instruments for surgery.
- Product code
- GCJ Laparoscope, General & Plastic Surgery
- Reason for recall
- Applied Medical is conducting a coluntary recall of two lot numbers, 1084426 and 1085506, of the 15mm Kii non-threaded optical access system, model C0R36, due to a potential weakness in the cannula tip, possibly resulting in tip breakage.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Applied Medical Resources Corp. notified Customers via an overnight "Voluntary Recall Notification" Letter dated June 9, 2009. The letter informed customers of the reason for recall and asked that the listed product be returned. The recall letter provided customers with a contact and telephone number for product return questions (1-949-713-8652) and a separate contact for regulatory questions (1-949-713-8369).
- Quantity in commerce
- 934 units
- Distribution
- Class II Recall - Worldwide -- United States, Australia, New Zealand, France, Canada, Norway and Great Britain.
- Product codes and lots
- Model Number: C0R36, Lot Numbers:1084426 and 1085506.
FIRST ARTICLE / WEEKLY