Ultra-Thin LightMat Surgical Illuminator, Disposable one time use. Model Number: UA2550, Exp Date: 2012-05...
Lumitex Inc began a recall of Ultra-Thin LightMat Surgical Illuminator, Disposable one time use. Model Number: UA2550, Exp Date: 2012-05. Product is packaged in individual sterile pouch w/5... on . The reason FDA records is "Adhesive which is used to attach LightMat to retractor may not have been tested to Medical standards. The vendor provided an adhesive strip that did not conform to Lumitex specifications. The adhesive strip provided may...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1998-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Lumitex Medical Devices, Inc.
- Product
- Ultra-Thin LightMat Surgical Illuminator, Disposable one time use. Model Number: UA2550, Exp Date: 2012-05. Product is packaged in individual sterile pouch w/5 pouches per box. The LightMat is intended for the illumination of surgical procedures, particularly where deep cavities or adjacent tissues limit outside light in the surgical field.
- Product code
- FDG Retractor, Fiberoptic
- Reason for recall
- Adhesive which is used to attach LightMat to retractor may not have been tested to Medical standards. The vendor provided an adhesive strip that did not conform to Lumitex specifications. The adhesive strip provided may not have met Lumitex requirements to be medical grade, EtO compatible, and biocompatible.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Lumitex, Inc. issued a recall letter dated July 2, 2009 to their Customers describing the affected device and requesting its return. For further information, contact Lumitex, Inc. at 1-440-243-8401.
- Quantity in commerce
- 59 packs/5 individual units
- Distribution
- Product was distributed to the following states: CA, FL, KS, MA, ME, MI , NC, NM, NV, NY, OH, and TX.
- Product codes and lots
- Model Number: UA2550; Lot Number: I01161.