Maxi Move Scoop Stretcher; Note: The stretcher is an accessory to the Maxi Move Patient Lift; Made by Arjo...
ArjoHuntleigh began a recall of Maxi Move Scoop Stretcher; Note: The stretcher is an accessory to the Maxi Move Patient Lift; Made by Arjo Hospital Equipment AB, Eslov, Sweden; product number... on . The reason FDA records is "A Scoop Stretcher is made up of two frames that are held together with two locking devices, one on each end. Each locking device contains a safety latch. The safety latch could unintentionally release when the side...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2230-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Arjohuntleigh AB
- Product
- Maxi Move Scoop Stretcher; Note: The stretcher is an accessory to the Maxi Move Patient Lift; Made by Arjo Hospital Equipment AB, Eslov, Sweden; product number KRA2210-20.
- Product code
- FSA Lift, Patient, Non-Ac-Powered
- Reason for recall
- A Scoop Stretcher is made up of two frames that are held together with two locking devices, one on each end. Each locking device contains a safety latch. The safety latch could unintentionally release when the side frames are pulled for positioning.
- Root cause tag
- manufacturing process
- FDA root cause
- Manufacturing material removal
- Action
- ArjoHuntleigh sent their customers a cover letter and the Field Safety Notification, informing them of the potential problem with the safety latches on the stretcher and a pending visit from an ArjoHuntleigh Service Technician to correct the problem. These letters were sent via UPS 2nd day delivery on 8/18/09. The customers were requested to test their stretchers upon receipt of the letter following the instructions for checking the safety latches on the stretcher: * Place the stretcher on its side on a non-scratch surface. * Try to pull the two sides apart with the safety latch engaged. * Test both the head and foot ends of the stretcher. * If the stretcher pulls apart, remove it from use immediately. It is important to note that all affected stretchers will be upgraded, whether they fail the test or not.
- Quantity in commerce
- 214 units
- Distribution
- Worldwide Distribution
- Product codes and lots
- all six digit serial numbers beginning with 6xxxxx and all serial numbers 08xxxxxxxx up to 0812001186.