ARCHITECT i2000 System and ARCHITECT i2000 SR wash buffer and wash zone mechanism grounding kit. ARCHITECT...
Abbott Laboratories, Inc began a recall of ARCHITECT i2000 System and ARCHITECT i2000 SR wash buffer and wash zone mechanism grounding kit. ARCHITECT System is designed to perform automated chemistry... on . The reason FDA records is "Under certain conditions, the product's wash buffer containing sodium azide can come in contact with the wash zone mechanism ground cables on either of the two wash zones. The copper in the ground strap may corrode and...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0132-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- ARCHITECT i2000 System and ARCHITECT i2000 SR wash buffer and wash zone mechanism grounding kit. ARCHITECT System is designed to perform automated chemistry tests utilizing photometry and potentiometric technology; and immunoassay tests utilizing chemiluminescent microparticle assay detection technology.
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- Under certain conditions, the product's wash buffer containing sodium azide can come in contact with the wash zone mechanism ground cables on either of the two wash zones. The copper in the ground strap may corrode and form an unstable chemical substance (metal azide) that may be sensitive to direct pressure and impact.
- FDA root cause
- Other
- Action
- A product correction notice was issued to consignees on July 17, 2009. Field representatives were instructed to visit consignees, inspect cables and replace them
- Quantity in commerce
- 4,526
- Distribution
- Worldwide Distribution
- Product codes and lots
- ARCHITECT i2000 System List Number 8C89-01; ARCHITECT i2000SR System System List Number 3M74-01 and 3M74-02; Wash Zone Mechanism Grounding Kit Part Number 7-201369-01; Concentrated Wash Buffer 1 L List Number 6C54-58 and l0 L List number 6C54-88.