Z-1693-2009

Heater Elements for Resuscitaire Infant Radiant Warmers model numbers RW82, RW82 VHA, WBR82, and WMRW82, 120...

Draeger Medical Systems, Inc. / initiated 2009-07-29 / Terminated / root cause: manufacturing process

Draeger Medical Systems, Inc. began a recall of Heater Elements for Resuscitaire Infant Radiant Warmers model numbers RW82, RW82 VHA, WBR82, and WMRW82, 120 volt configuration only on . The reason FDA records is "Spark, flame and smoke: A mis-positioned crimp on heater element lead wire resulted in flame and smoke aerating from the top of the warmer assembly". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1693-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Draeger Medical Systems, Inc.
Product
Heater Elements for Resuscitaire Infant Radiant Warmers model numbers RW82, RW82 VHA, WBR82, and WMRW82, 120 volt configuration only.
Product code
FMT Warmer, Infant Radiant
Reason for recall
Spark, flame and smoke: A mis-positioned crimp on heater element lead wire resulted in flame and smoke aerating from the top of the warmer assembly.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Draeger Medical issued an Urgent Medical Device Recall letter via certified mail on 8/7/09. The letter explained the problem, the need to conduct an inspection of the infant warmer (instructions included), and request replacement elements.
Quantity in commerce
2915 elements
Distribution
Worldwide distribution: USA, Canada, Saudi Arabia, Bermuda, Virgin Islands, Mexico, Panama, Colombia, and Costa Rica.
Product codes and lots
Model numbers RW82, RW82 VHA, WBR82, and WMRW82. Date codes ending in 07