Heater Elements for Resuscitaire Infant Radiant Warmers model numbers RW82, RW82 VHA, WBR82, and WMRW82, 120...
Draeger Medical Systems, Inc. began a recall of Heater Elements for Resuscitaire Infant Radiant Warmers model numbers RW82, RW82 VHA, WBR82, and WMRW82, 120 volt configuration only on . The reason FDA records is "Spark, flame and smoke: A mis-positioned crimp on heater element lead wire resulted in flame and smoke aerating from the top of the warmer assembly". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1693-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Draeger Medical Systems, Inc.
- Product
- Heater Elements for Resuscitaire Infant Radiant Warmers model numbers RW82, RW82 VHA, WBR82, and WMRW82, 120 volt configuration only.
- Product code
- FMT Warmer, Infant Radiant
- Reason for recall
- Spark, flame and smoke: A mis-positioned crimp on heater element lead wire resulted in flame and smoke aerating from the top of the warmer assembly.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Draeger Medical issued an Urgent Medical Device Recall letter via certified mail on 8/7/09. The letter explained the problem, the need to conduct an inspection of the infant warmer (instructions included), and request replacement elements.
- Quantity in commerce
- 2915 elements
- Distribution
- Worldwide distribution: USA, Canada, Saudi Arabia, Bermuda, Virgin Islands, Mexico, Panama, Colombia, and Costa Rica.
- Product codes and lots
- Model numbers RW82, RW82 VHA, WBR82, and WMRW82. Date codes ending in 07