Z-0198-2010
Zimmer VerSys Hip System Femoral Head, 12/14 taper, 28 mm dia., neck length +0 mm, sterile, Zimmer, Warsaw...
Zimmer GmbH / initiated 2009-08-03 / Terminated / root cause: manufacturing process
Zimmer Inc. began a recall of Zimmer VerSys Hip System Femoral Head, 12/14 taper, 28 mm dia., neck length +0 mm, sterile, Zimmer, Warsaw, IN; REF . Intended for mating femoral... on . The reason FDA records is "Manufacturing material may be present on the surface of the implant". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0198-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Zimmer VerSys Hip System Femoral Head, 12/14 taper, 28 mm dia., neck length +0 mm, sterile, Zimmer, Warsaw, IN; REF 8018-28-02. Intended for mating femoral stems equipped with tapered necks.
- Product code
- LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
- Reason for recall
- Manufacturing material may be present on the surface of the implant.
- Root cause tag
- manufacturing process
- FDA root cause
- Manufacturing material removal
- Action
- An Urgent: Device Recall letter, dated 8/3/2009, was issued to consignees. The letter identified the affected product, stated the reason for the recall, the clinical implications (risks to health), and actions to be taken. Customers are to locate any unused devices and quarantine them immediately, carry out a physical count of affected product and record the data on the Return Certification Form included with the letter, and fax a copy of the completed form back to Zimmer, Inc. at 574-372-4265. Customers are to return the recalled product along with the completed Inventory Return Certification Form. A "Dear Surgeon" letter will be issued to any hospitals where the devices have been implanted.
- Quantity in commerce
- 104
- Distribution
- Worldwide Distribution -- USA, Brazil, Canada, and Switzerland.
- Product codes and lots
- Lot 60378581.