Pointe Scientific Direct Bilirubin Reagent Set, Pointe Scientific Inc., Canton, MI., Catalog # HB936-1250
Pointe Scientific, Inc. began a recall of Pointe Scientific Direct Bilirubin Reagent Set, Pointe Scientific Inc., Canton, MI., Catalog # HB936-1250 on . The reason FDA records is "R2 reagent component may fail because it may contain R1 reagent. The reagent was rendered inoperative and would not pass the required quality control (calibration and quality control testing) before beginning running...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1014-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Horiba Instruments Incorporated
- Product
- Pointe Scientific Direct Bilirubin Reagent Set, Pointe Scientific Inc., Canton, MI., Catalog # HB936-1250.
- Product code
- CIG Diazo Colorimetry, Bilirubin
- Reason for recall
- R2 reagent component may fail because it may contain R1 reagent. The reagent was rendered inoperative and would not pass the required quality control (calibration and quality control testing) before beginning running the bilirubin assay.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- Consignee was notified by letter dated 7/29/2009 and instructed to identify and notify them at once of this product recall. Product is to be returned or destroyed. Customers with questions should contact the firm directly.
- Quantity in commerce
- 2
- Distribution
- California.
- Product codes and lots
- Lot 817103-337.