Medtronic Virtuoso DR, Model D154AWG, Dual Chamber implantable cardioverter defibrillator with atrial and...
Medtronic Inc. Cardiac Rhythm Managment began a recall of Medtronic Virtuoso DR, Model D154AWG, Dual Chamber implantable cardioverter defibrillator with atrial and ventricular therapies. Medtronic, Inc., 710 Medtronic... on . The reason FDA records is "Medtronic has detected a specific pattern of MOSFET IC malfunctions in its Concerto, Virtuoso and EnRhythm family of devices. The probability of occurrence decreases with time and, to date over 90% of the malfunctions...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1953-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Medtronic Virtuoso DR, Model D154AWG, Dual Chamber implantable cardioverter defibrillator with atrial and ventricular therapies. Medtronic, Inc., 710 Medtronic Parkway, Minneapolis, MN 55432-5604 USA. This device is indicated to provide ventricular antitachycardia pacing and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias in patients with NYHA functional Class II/III heart failure. In addition, the device is indicated for use in the above patients with atrial tachyarrhythmias, or those patients who are at significant risk of developing atrial tachyarrhythmias.
- Product code
- LWS Implantable Cardioverter Defibrillator (Non-Crt)
- Reason for recall
- Medtronic has detected a specific pattern of MOSFET IC malfunctions in its Concerto, Virtuoso and EnRhythm family of devices. The probability of occurrence decreases with time and, to date over 90% of the malfunctions related to the pattern have occurred within the first twelve months after implant.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Medtronic, Inc. personnel retrieved 2 non-implanted devices remaining at 2 hospitals. A Performance Notes letter was left at each facility. The Performance Notes described the issue, actions to address issue and recommendations to physicians. The Performance Notes will also be posted on the firm's website. Direct questions about this recall to Medtronic, Inc. by calling 1-763-514-4000.
- Quantity in commerce
- 1
- Distribution
- Nationwide Distribution-- (states of IA and NJ).
- Product codes and lots
- Serial number: PUL450634H.