Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2019-2009

Bard * Button Device Continuous Feeding Tube with 90 degree Adaptor, 18F x 1.7cm, NON STERILE, REF 000282...

Bard Access Systems, Inc. / initiated 2009-01-20 / Terminated / root cause: manufacturing process

C R Bard Inc began a recall of Bard * Button Device Continuous Feeding Tube with 90 degree Adaptor, 18F x 1.7cm, NON STERILE, REF 000282, Bard Access Systems, Inc., Salt Lake City, UT 84116... on . The reason FDA records is "Possible restriction/occlusion in adaptors/tube accessories causing difficulty to infuse during routine feeding treatment". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2019-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bard Access Systems, Inc.
Product
Bard * Button Device Continuous Feeding Tube with 90 degree Adaptor, 18F x 1.7cm, NON STERILE, REF 000282, Bard Access Systems, Inc., Salt Lake City, UT 84116 USA. The device is designed for use with the BARD Button Replacement Gastrostomy Device and used for enteral feeding in patients requiring long-term nutritional support.
Product code
KGC Tube, Gastro-Enterostomy
Reason for recall
Possible restriction/occlusion in adaptors/tube accessories causing difficulty to infuse during routine feeding treatment.
Root cause tag
manufacturing process
FDA root cause
Equipment maintenance
Action
Customers were notified by letter and instructed to not use or further distribute any of the affected lot numbers. Customers were further instructed to return the recalled products to Bard Access Systems (BAS) through a return goods authorization number. Questions and requests for assistance should be directed to BAS Customer Service at 800-290-1689.
Quantity in commerce
13, 337 total of all products
Distribution
Worldwide Distribution -- US, Canada, Austria, Australia, Belgium, Finland, France, Germany, Great Britain, Holland, Italy, Norway, Spain, Sweden and Switzerland, Japan, and Latin America.
Product codes and lots
Lot Numbers: HUSD1960 and HUSD1959.