Bard * Button Device Continuous Feeding Tube with 90 degree Adaptor, 18F x 1.7cm, NON STERILE, REF 000284...
C R Bard Inc began a recall of Bard * Button Device Continuous Feeding Tube with 90 degree Adaptor, 18F x 1.7cm, NON STERILE, REF 000284, Bard Access Systems, Inc., Salt Lake City, UT 84116... on . The reason FDA records is "Possible restriction/occlusion in adaptors/tube accessories causing difficulty to infuse during routine feeding treatment". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2021-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Bard Access Systems, Inc.
- Product
- Bard * Button Device Continuous Feeding Tube with 90 degree Adaptor, 18F x 1.7cm, NON STERILE, REF 000284, Bard Access Systems, Inc., Salt Lake City, UT 84116 USA. The device is designed for use with the BARD Button Replacement Gastrostomy Device and used for enteral feeding in patients requiring long-term nutritional support.
- Product code
- KGC Tube, Gastro-Enterostomy
- Reason for recall
- Possible restriction/occlusion in adaptors/tube accessories causing difficulty to infuse during routine feeding treatment.
- Root cause tag
- manufacturing process
- FDA root cause
- Equipment maintenance
- Action
- Customers were notified by letter and instructed to not use or further distribute any of the affected lot numbers. Customers were further instructed to return the recalled products to Bard Access Systems (BAS) through a return goods authorization number. Questions and requests for assistance should be directed to BAS Customer Service at 800-290-1689.
- Quantity in commerce
- 13, 337 total of all products
- Distribution
- Worldwide Distribution -- US, Canada, Austria, Australia, Belgium, Finland, France, Germany, Great Britain, Holland, Italy, Norway, Spain, Sweden and Switzerland, Japan, and Latin America.
- Product codes and lots
- Lot Numbers: HUSD1611, HUSD1963, HUSD0119, HUSD0120 and HUSD1962.