Z-0554-2010

Zimmer / Abbott Spine Sequoia dorsal height & revision tool, non-sterile; REF 3367-1. Surgical instrument for...

Zimmer GmbH / initiated 2009-08-20 / Terminated / root cause: design

Zimmer Inc. began a recall of Zimmer / Abbott Spine Sequoia dorsal height & revision tool, non-sterile; REF 3367-1. Surgical instrument for pedicle screw system on . The reason FDA records is "The tip of the instrument may fracture during use". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0554-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Zimmer / Abbott Spine Sequoia dorsal height & revision tool, non-sterile; REF 3367-1. Surgical instrument for pedicle screw system.
Product code
NKB Thoracolumbosacral Pedicle Screw System
Reason for recall
The tip of the instrument may fracture during use.
Root cause tag
design
FDA root cause
Device Design
Action
Zimmer sent a letter dated 8/17/09 to distributors and consignees on or about 8/20/09. Users were notified of the potential problem via the letter and advised that the instruments will be removed and replaced when a new version of the instrument becomes available (around 10/09).
Quantity in commerce
215
Distribution
Nationwide, Belgium and New Zealand.
Product codes and lots
Lots 47PN, 47PP, 49HK, 49HL and 57WL.