Tornier, LMP-50 (Lesser Metatarsal Phalangeal Implant), Size 50, Sterile EO, , Manufactured by Tornier, Inc...
Tornier, Inc. began a recall of Tornier, LMP-50 (Lesser Metatarsal Phalangeal Implant), Size 50, Sterile EO, , Manufactured by Tornier, Inc, Edina, MN 55435. The Lesser Metatarsal Phalangeal... on . The reason FDA records is "Tornier, Inc. has received a report of a mislabeled product due to a product mix-up. The package label indicated that the product was a model LMP-50 however the product inside the package was Tornier FGT-50 implant...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0455-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- TORNIER Inc.
- Product
- Tornier, LMP-50 (Lesser Metatarsal Phalangeal Implant), Size 50, Sterile EO, , Manufactured by Tornier, Inc, Edina, MN 55435. The Lesser Metatarsal Phalangeal Joint Implant is indicated for: Partial or complete dislocation of the lesser metatarsophalangeal joint.
- Product code
- KWH Prosthesis, Toe, Constrained, Polymer
- Reason for recall
- Tornier, Inc. has received a report of a mislabeled product due to a product mix-up. The package label indicated that the product was a model LMP-50 however the product inside the package was Tornier FGT-50 implant (Primus Flexible Great Toe Implant With Grommetts).
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling mix-ups
- Action
- Consignees were sent on April 6, 2009 a Tornier "Urgent-Product Recall" letter dated April 3, 2009. The letter was addressed to Healthcare professionals and distributors. The letter described the problem and product involved. Advised consignees to immediately discontinue use of the affected product and to return any unused affected product to their facility. Advised their distributors to notify all accounts. They also attached a questionnaire needed for inventory status.
- Quantity in commerce
- 77
- Distribution
- Worldwide Distribution -- USA, including states of CA, FL, MD, MI, NJ, OH, and PA, and country of France.
- Product codes and lots
- Lot number: K10747, K10677, and K10510.