Z-1017-2010
Dimension Vista System BUN Flex Reagent Cartridge, Catalog number K1021, in vitro diagnostic
Siemens Healthcare Diagnostics Inc. / initiated 2009-06-16 / Terminated / root cause: manufacturing process
Siemens Healthcare Diagnostics, Inc. began a recall of Dimension Vista System BUN Flex Reagent Cartridge, Catalog number K1021, in vitro diagnostic on . The reason FDA records is "Incorrect result: patient sample BUN results reported may be higher or lower than actual level". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1017-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Dimension Vista System BUN Flex Reagent Cartridge, Catalog number K1021, in vitro diagnostic.
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- Incorrect result: patient sample BUN results reported may be higher or lower than actual level.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Siemens issued an Urgent Field Safety Notice dated 6/16/09 to all affected customers instructing them to discontinue use of the Dimension Vista BUN Flex reagent cartridge lot 09061AA.
- Quantity in commerce
- 1254 cartons
- Distribution
- Worldwide distribution: USA, Australia, Canada, Germany, Japan, and Korea.
- Product codes and lots
- Lot number 09061AA exp 2010-03-02