Oridion Medical Capnostream20 capnograph/pulse oximeter monitor and monitor with printer. Part numbers CS...
Oridion Medical 1987 Ltd. began a recall of Oridion Medical Capnostream20 capnograph/pulse oximeter monitor and monitor with printer. Part numbers CS 08651, CS 08653, CS08657, and CS08659 on . The reason FDA records is "1) The monitor's ability to withstand electrostatic discharge (ESD) to exposed metal connectors on the back of the monitor which may cause the monitor's display to freeze. 2) The unit generates electromagnetic radiation...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2232-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Oridion Medical 1987 Ltd.
- Product
- Oridion Medical Capnostream20 capnograph/pulse oximeter monitor and monitor with printer. Part numbers CS 08651, CS 08653, CS08657, and CS08659.
- Product code
- CCK Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase
- Reason for recall
- 1) The monitor's ability to withstand electrostatic discharge (ESD) to exposed metal connectors on the back of the monitor which may cause the monitor's display to freeze. 2) The unit generates electromagnetic radiation above the limits set by applicable standards for medical devices
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Oridian sent a letter to all its distributors June 25, 2009, instructing them to gather all the units to be exchanged for factory reconditioned units that have the hardware and software enhancements and meet all performance requirements in full.
- Quantity in commerce
- 326 units
- Distribution
- Nationwide.
- Product codes and lots
- Software version V 3.0 and earlier. Part Number Part Description CS08651, Capnostream20 with Nellcor pulse oximetry module CS08653 Capnostream20 with Masimo pulse oximetry module CS08657 Capnostream20 with Nellcor pulse oximetry module and printer CS08659 Capnostream20 with Masimo pulse oximetry module and printer