Bard Access Systems, Inc., FASTRAC* "Pull" Gastric Access Port Safety System, REF 006033, 20F, Sterile
C.R. Bard Access Systems, Inc began a recall of Bard Access Systems, Inc., FASTRAC* "Pull" Gastric Access Port Safety System, REF 006033, 20F, Sterile on . The reason FDA records is "Snare can detach in patient. The Polypectomy Snares and the Entake Peg System (which contain these snares) are being recalled because the looped portion of the snare can become detached and drop into the stomach or...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2185-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Bard Access Systems, Inc.
- Product
- Bard Access Systems, Inc., FASTRAC* "Pull" Gastric Access Port Safety System, REF 006033, 20F, Sterile.
- Product code
- KNT Tubes, Gastrointestinal (And Accessories)
- Reason for recall
- Snare can detach in patient. The Polypectomy Snares and the Entake Peg System (which contain these snares) are being recalled because the looped portion of the snare can become detached and drop into the stomach or other internal canal during use, necessitating retrieval of the remaining piece.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Bard contacted all 1,883 worldwide customers who received the distributed suspect 467 product code lots, comprising 177, 614 units that were either not expired or were still available in the field. These will be returned by the customer to Bard Access Systems. 1-800-290-1689.
- Quantity in commerce
- 177,614 units
- Distribution
- Worldwide, including USA, Belgium, Canada, Australia, and Japan.
- Product codes and lots
- Lots: 43ARA073, 43DQA028, 43DQA029, 43DQA030, 43FQA053, 43FOA054, 43FQA055, 43FQA056, 43FQA057, 43FQA058, 43FQA059, 43GQA058, 43GQA059, 43GQA060, 43GQA061, 43GQA062, 43GQA063, 43GQA064, 43GQA065, 43HQA053, 43HQA187, 43IQA054, 43JQA055, 43JQA056, 43KQA070, 43KQA071, 43KQA072, 43LQA073, 43LQA074, HURA1470, HURA1471, HURA1472, HURA1473, HURB2558, HURB2559, HURD1498, HURD1644, HURD1798, HURD1799, HURD1800, HURD1801, HURE2820, HURE2821, HURE2822, HURE3300, HURE3301, HURE3302, HURE3303, HURF0404, HURF0405, HURF0406, HURG0210, HURG1382, HURG1383, HURG1384, HURG1385, HURH0869, HURH0870, HURH0871, HURH0872, HURH0873, HURI0518, HURI1003, HURI1203, HURI1204, HURI1205, HURJ0306, HURJ0308, HURJ0309, HURJ0312, HURJ1440, HURK1457, HURK1458, HURK1459, HURK1460, HUSA0275, HUSA0276, HUSB0097, HUSC0158, HUSC0159, HUSC1201, HUSC1763, HUSD0095, HUSD0096, HUSD0097, HUSD0098, HUSD1936, HUSD1937, HUSD1938, HUSD1939, HUSE1505, HUSE1506, HUSE1508, HUSF0001, HUSF1748.