ACUSON Sequoia 12.X C512/512 SS Base System, model number 10042700. Distributed by Siemens Medical Solutions...
Siemens Medical Solutions USA, Inc. began a recall of ACUSON Sequoia 12.X C512/512 SS Base System, model number 10042700. Distributed by Siemens Medical Solutions USA, Inc., Business Unit Ultrasound, Mountain... on . The reason FDA records is "Shock Hazard. The manufacturer failed to set the system power switch to "ON" during the line level leakage and high pot current draw testing". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2116-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- ACUSON Sequoia 12.X C512/512 SS Base System, model number 10042700. Distributed by Siemens Medical Solutions USA, Inc., Business Unit Ultrasound, Mountain View, CA. The ACUSON Sequoia is a general purpose Diagnostic Ultrasound system. Its function is to acquire primary or secondary harmonic ultrasound echo data and display it in: B-Mode, M-Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Amplitude Doppler Mode, a combination of modes, Harmonic Imaging or 3D imaging on a FPD Display. The ACUSON Sequoia is intended for the following applications: Abdominal, General Imaging and Cardiolgoy for Fetal, Abdominal, Intraoperative (abdominal and neurological), Pediatrics, Small Organs (breat, testes, thyroid and penis), Neonatal/Adult Cephalic, Cardiac (adult, pediatric and neonatal), Transesophageal, Transrectal, Transvaginal, Peripheral Vessels and Musculo-sceletal (superficial and conventional) applications. The system also provides for the measurement of anatomical structures that provide information that is used for clinical diagnosis purposes.
- Product code
- IYN System, Imaging, Pulsed Doppler, Ultrasonic
- Reason for recall
- Shock Hazard. The manufacturer failed to set the system power switch to "ON" during the line level leakage and high pot current draw testing.
- Root cause tag
- manufacturing process
- FDA root cause
- Release of Material/Component prior to receiving test results
- Action
- Siemens Medical Solutions USA, Inc. sent a Customer Safety Advisory Notification letter out on May 15, 2009. Service engineers were to visit each customer to conduct necessary testing. For questions regarding this recall call 650-694-5993.
- Quantity in commerce
- 10 units
- Distribution
- Worldwide Distribution - USA including CA, IL, MD, NC, OH, and UT and the countries of Australia, China, Denmark, Spain, South Korea, Norway, Poland, Russia and Sweden.
- Product codes and lots
- Serial numbers: 69275, 69301, 69313, 69335, 69347, 69353, 69366, 69391, 69403, and 69411.