CellSearch Circulating Tumor Cell Kit (Epithelial) IVD REF 7900001 Veridex LLC, a Johnson & Johnson Company
Veridex, LLC began a recall of CellSearch Circulating Tumor Cell Kit (Epithelial) IVD REF 7900001 Veridex LLC, a Johnson & Johnson Company on . The reason FDA records is "Veridex has received 4 complaints regarding reagent bottles having been incorrectly placed or duplicate bottles places within the reagent tray of CellSearch Circulating Tumor Cell Kit. Delayed detection of disease...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0819-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Veridex, LLC
- Product
- CellSearch Circulating Tumor Cell Kit (Epithelial) IVD REF 7900001 Veridex LLC, a Johnson & Johnson Company.
- Product code
- NQI System, Immunomagnetic, Circulating Cancer Cell, Enumeration
- Reason for recall
- Veridex has received 4 complaints regarding reagent bottles having been incorrectly placed or duplicate bottles places within the reagent tray of CellSearch Circulating Tumor Cell Kit. Delayed detection of disease relapse or progression may occur due to false low CTC results.
- Root cause tag
- packaging
- FDA root cause
- Other
- Action
- Urgent Product Correction Notification letters were sent to all customers on August 12, 2009 by Federal Express. Customers were advised to inspect the kits to assure the proper reagents are in the proper position. If not, do not use the kits and contact the company at 877-837-4339 (chose option 1).
- Quantity in commerce
- 1641 in US, 380 internationally.
- Distribution
- Product was distributed worldwide, US (one government account: National Cancer Institute, Frederick, MD). Also 8 international accounts in Belgium, France, Germany, Italy, Spain, United Kingdom, Netherlands, Norway and Israel.
- Product codes and lots
- Lot number 0019, Exp 1/8/2010; 0019B, Exp 1/8/2010; 0039, Exp 3/16/2010; 0059, Exp 3/6/2010, 0079, Exp 2/20/2010; 0099, Exp 6/24/2010.