Z-1404-2010

Renessa RF System, model PR0918, manufactured by Novasys Medical, Newark, CA

Novasys Medical, Inc. / initiated 2009-07-27 / Terminated / root cause: labeling and UDI

Novasys Medical Inc began a recall of Renessa RF System, model PR0918, manufactured by Novasys Medical, Newark, CA on . The reason FDA records is "Instructions for Use (IFU) revision-- Because of complaints received by the firm of worsened stress incontinence symptoms 3-4 weeks after treatment, the instructions for use were revised to emphasize potential side...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1404-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Novasys Medical, Inc.
Product
Renessa RF System, model PR0918, manufactured by Novasys Medical, Newark, CA
Product code
NVJ Applicator, Transurethral, Radio Frequency, For Stress Urinary Incontinence In Women
Reason for recall
Instructions for Use (IFU) revision-- Because of complaints received by the firm of worsened stress incontinence symptoms 3-4 weeks after treatment, the instructions for use were revised to emphasize potential side effects. Although Novasys has not, to date, identified a device malfunction associated with the symptoms discussed above, the investigation over time shows that there is a correlation
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
A Field Action notification letter was sent to all current customers (physicians) along with the revised instructions for use, and a small supply of the revised patient information brochures. The company also updated physician information brochures and patient/physician DVDs to include the information. This was shipped to physicians in August/September of 2009. Additional copies of the Patient Brochures can be ordered by contacting Novasys Customer Service at 866-784-4777. Questions should be addressed to the firm at the same telephone number.
Quantity in commerce
As of 6/12/09 there were 436 Probes at customer accounts.
Distribution
US nationwide only. (TX, FL, CA, LA, KY, MO, GA,OR WA, HI, AZ, OK AL, TN) .
Product codes and lots
Issue affects all lots in distribution.