Z-0064-2010

200996-Cable Assembly, Hybrid Extended Length, 200997 Cable Assembly, Hybrid, Guidance Module, 200998 Cable...

Mako Surgical Corp. / initiated 2009-07-17 / Terminated / root cause: design

Mako Surgical Corporation began a recall of 200996-Cable Assembly, Hybrid Extended Length, 200997 Cable Assembly, Hybrid, Guidance Module, 200998 Cable Assembly, Hybrid, UI Panel Subcomponents for the... on . The reason FDA records is "Cables fail to provide adequate transmission. MAKO Surgical Corporation is recalling 200996 Cable Assembly, Hybrid Extended Length; 200997 Cable Assembly, Hybrid, Guidance Module; and 200998 Cable Assembly, Hybrid, UI...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0064-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Mako Surgical Corp.
Product
200996-Cable Assembly, Hybrid Extended Length, 200997 Cable Assembly, Hybrid, Guidance Module, 200998 Cable Assembly, Hybrid, UI Panel Subcomponents for the 203000 RIO STANDARD SYSTEM Manufactured by: MAKO Surgical, Corp. Fort Lauderdale, FL. All three cables are integral components and are not purchased separately. However, 200996-Cable Assembly is the only cable visible. This cable is labeled: 200996-REV-1, KSM 390634-1, SN 000001. 200997 and 200998 Cable Assembly are not visible . The Tactile Guidance System v2.0 is intended to assist the surgeon in providing software defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures.
Product code
HAW Neurological Stereotaxic Instrument
Reason for recall
Cables fail to provide adequate transmission. MAKO Surgical Corporation is recalling 200996 Cable Assembly, Hybrid Extended Length; 200997 Cable Assembly, Hybrid, Guidance Module; and 200998 Cable Assembly, Hybrid, UI Panel. These parts are sub components of the 20300 Rio Standard System.
Root cause tag
design
FDA root cause
Device Design
Action
A Recall Notice, dated July 15, 2009, was sent to owners of the RIO system (hospital OR staff) with a signed acknowledgement returned to MAKO. Surgeons using the RIO system during the recall time period, regardless of whether their system has already been upgraded, are given the notification letter and asked to sign and return the acknowledgement memo. The letter stated that customers' RIO will be upgraded to replace the potentially faulty cable with a new cable. Their MAKOplasty Specialist (MPS) will notify them when their system was scheduled to have their fiber optic cable upgrade. The upgrade will be installed by a Field Service Engineer (FSE) free of charge and the RIO will be completely tested prior to being returned to service. Customers should direct questions or concerns to their MPS, FSE, or Jim Pomeroy, who can be reached at 954-927-2044.
Quantity in commerce
23
Distribution
Nationwide Distribution
Product codes and lots
Serial Numbers: ROB031 through ROB068.