Philips Medical Systems North America Co. Phillips began a recall of OmniDiagnost on . The reason FDA records is "The x-ray system fail to meet the requirements of 21 CFR 1020.32(d)(2)(ii) when a specific tube failure occurs involving fluoroscopy of large patients and a partial failure of the grid switch". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1665-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- OmniDiagnost
- Product code
- JAA System, X-Ray, Fluoroscopic, Image-Intensified
- Reason for recall
- The x-ray system fail to meet the requirements of 21 CFR 1020.32(d)(2)(ii) when a specific tube failure occurs involving fluoroscopy of large patients and a partial failure of the grid switch
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- A Customer Notification letter describing the potential grid switch failure and actions planned by Philips to correct the problem will be sent to all users of the affected systems during the week of July 6, 2009.
- Quantity in commerce
- 28 units
- Distribution
- Worldwide Distribution
- Product codes and lots
- N/A