Medtronic Physio-Control LIFEPAK CR Plus Defibrillator/Monitor. Automated external defibrillator deployed...
Physio Control, Inc. began a recall of Medtronic Physio-Control LIFEPAK CR Plus Defibrillator/Monitor. Automated external defibrillator deployed primarily in public access defibrillator environments... on . The reason FDA records is "The K1 relay on the Analog printed circuit board assembly may short due to moisture ingress in a highly humid environment. K1 shorting may affect the ECG amplitude causing the Shock Advisory System(SAS) to not analyze...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2030-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Physio Control, Inc.
- Product
- Medtronic Physio-Control LIFEPAK CR Plus Defibrillator/Monitor. Automated external defibrillator deployed primarily in public access defibrillator environments where the expected use for any one device is once every 20 months. For use on persons requiring defibrillation therapy.
- Product code
- MKJ Automated External Defibrillators (Non-Wearable)
- Reason for recall
- The K1 relay on the Analog printed circuit board assembly may short due to moisture ingress in a highly humid environment. K1 shorting may affect the ECG amplitude causing the Shock Advisory System(SAS) to not analyze the ECG rhythm correctly thus preventing the device from delivering therapy.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Physio-Control, Inc. notified Customers by phone beginning August 18, 2009. In addition, the recalling firm Issued an "Urgent Medical Device Recall" letter dated August 2009 informing Customers of the affected product and informing them that a Physio-Control representative will call and arrange for immediate device replacement. For further information, contact Physio-Control Customer Care at 1-800-442-1142.
- Quantity in commerce
- 34 devices (2 domestically & 32 internationally)
- Distribution
- Worldwide Distribution -- US (DC and MN) and Japan.
- Product codes and lots
- Serial Numbers: 37026963, 37026983, 37026984, 37026997, 37027002, 37027008, 37027039, 37027040, 37027049, 37027053, 37027063, 37027065, 37027066, 37027070, 37027071, 37027073, 37027075, 37027090, 37027099, 37027105, 37027122, 37027197, 37027529, 37027569, 37031393, 37037850, 37037893, 37037986, 37038002, 37038211, 37038365, 37135986, 37154526 and 37154638.