Z-0050-2012

Access Immunoassay Systems Total hCG. The Access Total hCG assay is a paramagnetic particle, chemiluminescent...

Siemens Healthcare Diagnostics Inc. / initiated 2009-03-05 / Terminated / root cause: manufacturing process

Beckman Coulter Inc. began a recall of Access Immunoassay Systems Total hCG. The Access Total hCG assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of... on . The reason FDA records is "The recall was initiated because Beckman Coulter has confirmed that certain Access Total hCG reagent packs show decreased stability from the date of manufacture, which may lead to increased imprecision". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0050-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Access Immunoassay Systems Total hCG. The Access Total hCG assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of total hCG levels in human serum and plasma using the Access Immunoassay Systems.
Product code
JHI Visual, Pregnancy Hcg, Prescription Use
Reason for recall
The recall was initiated because Beckman Coulter has confirmed that certain Access Total hCG reagent packs show decreased stability from the date of manufacture, which may lead to increased imprecision.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The recall communication was initiated on March 18, 2009 with Beckman Coulter forwarding an Urgent: Product Corrective Action (PCA) letter dated March 13, 2009 with attached PCA Response Form (via US Postal Service for US Customers) to all customers who purchased the Access Immunoassay Systems total hCG. The letter provides the customers with an explanation of the problem identified and an action to be taken. Customers were instructed to discontinue use of the reagent lot numbers 824542 and 825652, and discard any unused reagent packs. Customers were asked to review their historical records for low level quality control to ensure that the assay performance for the referenced lot numbers has been within their laboratory specifications. They were also asked to share the information with their laboratory staff and retain the letter for their Quality System documentation, and then complete and return the enclosed response form within 10 days so that they may be assured that consignees received the notification. Finally they were told to contact Customer Technical Support Center at 1-800-854-3633 in the US and Canada, or their local Beckman Coulter Representative with any questions.
Quantity in commerce
14,764
Distribution
Worldwide Distribution -- US, Algeria, Angola, Argentina, Armenia, Australia, Bangladesh, Belgium, Brazil, Bulgaria, canada, Chile, China, Colombia, Czech Republic, Egypt, El Salvador, Ethiopia, Finland, France, French Polynesia, Germany, Ghana, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Kenya, Korea, Kuwait, Lebanon, Libyan Arab Jamahiriya, Lithuania, Macao, Mexico, Monaco, Morocco, Namibia, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Philippines, Poland, Puerto Rico, Qatar, Reunion, Russian Federation, Saudi Arabia, Serbia, Singapore, South Africa, Spain, Swaziland, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Zambia.
Product codes and lots
Lot Numbers 824542 and 825652.