Z-0853-2011
RoboCouch Patient Support System, model number 025007, a subsystem of the Cyberknife Robotic Radiosurgery...
Accuray Incorporated / initiated 2009-07-21 / Terminated / root cause: design
Accuray Inc began a recall of RoboCouch Patient Support System, model number 025007, a subsystem of the Cyberknife Robotic Radiosurgery System, manufactured by Accuray Incorporated... on . The reason FDA records is "Mechanical Failure-- Patient Support System's table top may unexpectedly descend several inches". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0853-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Accuray Incorporated
- Product
- RoboCouch Patient Support System, model number 025007, a subsystem of the Cyberknife Robotic Radiosurgery System, manufactured by Accuray Incorporated, Sunnyvale, CA. Intended for use in the support and positioning of a patient during Radiosurgery and radiotherapy procedures and other medical procedures when precise positioning is required.
- Product code
- JAI Couch, Radiation Therapy, Powered
- Reason for recall
- Mechanical Failure-- Patient Support System's table top may unexpectedly descend several inches.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Consignees were notified verbally about the issue when an inspection was scheduled.
- Quantity in commerce
- 26
- Distribution
- Worldwide Distribution -- USA, including states of CA, FL, VA, AL, CT, MA, NC, TN, PA, RI, and NE and countries of Turkey, France, UK, Saudi Arabia, and Switzerland.
- Product codes and lots
- RoboCouch model 025007