Z-2061-2011

Siemens ACUSON X300 Premium Edition Systems with software versions 5.0.00 and 5.0.01 (model numbers 10348531...

Siemens Medical Solutions USA, Inc. / initiated 2009-07-24 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc. began a recall of Siemens ACUSON X300 Premium Edition Systems with software versions 5.0.00 and 5.0.01 (model numbers 10348531, 10348532 and 10348533), when using the V5Ms... on . The reason FDA records is "Product Failure-- When the 64-Channel System V5M is used in the 128-Channel mode, an over current fault occurs which shuts the system down". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2061-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens ACUSON X300 Premium Edition Systems with software versions 5.0.00 and 5.0.01 (model numbers 10348531, 10348532 and 10348533), when using the V5Ms transducer (model number 08263703) and the Plus option (10427795) Product is an ultrasound imaging system intended for General Radiology,Fetal, Abdominal, Intaoperatve, Pediatric, Small parts, Neunatal/Adult Cephalic, Cardiac, Transesophageal, Pelvic, Transcranial, OB/GYN, Urology, Vascular, Musculoskeletal, Superficial Musculoskeletal, and peripheral Vascular Applications. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes. The Arterial Health Package software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published with peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding the state of their cardiovascular system. The V5Ms is a multifrequency,multi-plane, phased sector array transducer, which can be introduced into the esophagus to obtain heart structure images. Transesophageal echocardiography is used to obtain unobstructed views of the heart surrounding tissue.
Product code
IYO System, Imaging, Pulsed Echo, Ultrasonic
Reason for recall
Product Failure-- When the 64-Channel System V5M is used in the 128-Channel mode, an over current fault occurs which shuts the system down.
Root cause tag
software
FDA root cause
Software design
Action
Siemens sent a Customer Safety Advisory Notification letter on July 24, 2009. The letter identified the product, the problem, and the action to be taken by the customer. A software update will be conducted in the field. Until that update occurs, customers were instructed to image without the CW doppler to avoid a system shut down. Customers were instructed to contact their service support person with any questions. For questions regarding this recall call 650-694-5993.
Quantity in commerce
70 units
Distribution
Worldwide Distribution - USA including CA, CT, FL, MA, and TX and the countries of United Arab Emirates, Germany, Ecuador, Spain, France, Great Britian, Saudi Arabia, Sweden, Thailand, and South Africa.
Product codes and lots
Premium Edition Systems with software versions 5.0.00 and 5.0.01 (model numbers 10348531, 10348532 and 10348533), when using the V5Ms transducer (model number 08263703) and the Plus option (10427795)