Foot Pedal, Foot Control for SC2000 and SC2100 Powered, simple cranial drills, burrs, trephines, and their...
The Anspach Effort, Inc. began a recall of Foot Pedal, Foot Control for SC2000 and SC2100 Powered, simple cranial drills, burrs, trephines, and their accessories. Catalog Numbers: EMAX2-FP and FPLUS-FP... on . The reason FDA records is "The Anspach Effort, Inc. is recalling the Foot Control/ Foot Pedal for SC2000 and SC2100 Powered, simple cranial drills, burrs, trephines, and accessories. The Foot Control Cables lacks the shield connection from the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0083-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- The Anspach Effort, Inc.
- Product
- Foot Pedal, Foot Control for SC2000 and SC2100 Powered, simple cranial drills, burrs, trephines, and their accessories. Catalog Numbers: EMAX2-FP and FPLUS-FP. Indicated to cut and shape bone including bones of the spine and cranium.
- Product code
- HBE Drills, Burrs, Trephines & Accessories (Simple, Powered)
- Reason for recall
- The Anspach Effort, Inc. is recalling the Foot Control/ Foot Pedal for SC2000 and SC2100 Powered, simple cranial drills, burrs, trephines, and accessories. The Foot Control Cables lacks the shield connection from the cable to the connector body. The lack of shielding may result in radiated emissions which could possibly cause interference on other electrical equipment adjacent to the foot pedal
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Anspach issued an "Urgent Product Removal" Letter dated August 17, 2009 informing customers of the affected product. Consignees are asked to screen their inventory and return any affected product including a completed Product Replacement Form to the firm. For further questions, contact Anspach at 1-800-327-6887 or 1-561-627-1080.
- Quantity in commerce
- 14
- Distribution
- Worldwide Distribution -- Canada, Indonesia, United Kingdom, Korea and United States (NC, OK, TN, CA, KS, MI and WA).
- Product codes and lots
- EMAX2-FP Serial Numbers: C26302250301, C26302250302, C26302250303, C26302250304, 5044044307, 31495607, 32380602, 31832209, 31832202 and 32366805; and EPLUS-FP Serial Numbers: C26302200804 , C26302200802, C27302221701 and C08301701901.