Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2065-2009

Portex Tracheal Tube, Oral/Nasal, Murphy Eye, Siliconized, 2.5 mm

Abbott Medical / initiated 2009-08-25 / Terminated / root cause: manufacturing process

Smiths Medical began a recall of Portex Tracheal Tube, Oral/Nasal, Murphy Eye, Siliconized, 2.5 mm Product Code: 100/141/025. Intended for oral and/or nasal intubation for airway management on . The reason FDA records is "Pediatric Trachel Tube manufactured with internal diameter smaller than indicated in the label may result in inability to remove secretions and cause partial or complete obstruction of the airway". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2065-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Portex Tracheal Tube, Oral/Nasal, Murphy Eye, Siliconized, 2.5 mm Product Code: 100/141/025. Intended for oral and/or nasal intubation for airway management.
Product code
BTR Tube, Tracheal (W/Wo Connector)
Reason for recall
Pediatric Trachel Tube manufactured with internal diameter smaller than indicated in the label may result in inability to remove secretions and cause partial or complete obstruction of the airway
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Smiths Medical issued an "Urgent Field Safety Notice" dated August 24, 2009 via UPS Second Day Mail including a Guidance Chart. Return of the enclosed Confirmation Form was also requested by fax to 1-603-355-8157. Smiths Medical issued a revised recall letter dated September 9, 2009. Customers are requested to return unused affected product to the firm. For further information, contact Smiths Medical ASD, Inc. by phone at 1-800-348-6064, prompt 4 extension 2273.
Quantity in commerce
60,325 (US) 322,560 (International)
Distribution
Worldwide Distribution -- United States, Andorra, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Barbados, Belgium, Bermuda, Borneo, Brazil, Brunei, Bulgaria, Canada, Chile, China, C¿¿te d'Ivoire, Czech Republic, Denmark, East Timor, Egypt, Estonia, Finland, France, Georgia, Germany, Gibraltar, Greece, Guatemala, Vatican City, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Lebanon, Libya, Liechtenstein, Lithuania, Luxembourg, Malaysia, Maldives, Malta, Mexico, New Zealand, Netherlands, Norway, Oman, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Senegal, Singapore, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Suriname, Sweden, Switzerland, Syria, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom and Uruguay.
Product codes and lots
Product manufactured before September 2009 and within its expiration date.