S&W Electrodes for Defibrillation, 10 Defibrillation Electrodes, Reorder #90101, S & W Healthcare Corp, 1525...
Leonhard Lang Gmbh began a recall of S&W Electrodes for Defibrillation, 10 Defibrillation Electrodes, Reorder #90101, S & W Healthcare Corp, 1525 Flight Path Drive Brooksville, Florida 34604 The... on . The reason FDA records is "A potential risk was identified associated with the connectors of the defibrillation electrodes DF20. It was found that some of the connectors do not slip easily into the sockets of the defibrillator or an extension...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2234-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Leonhard Lang GmbH
- Product
- S&W Electrodes for Defibrillation, 10 Defibrillation Electrodes, Reorder #90101, S & W Healthcare Corp, 1525 Flight Path Drive Brooksville, Florida 34604 The device is non-sterile and single use. For use on adults and children over eight years old for external defibrillation, pacing, monitoring, and cardioversion.
- Product code
- MKJ Automated External Defibrillators (Non-Wearable)
- Reason for recall
- A potential risk was identified associated with the connectors of the defibrillation electrodes DF20. It was found that some of the connectors do not slip easily into the sockets of the defibrillator or an extension cord of the defibrillator. In such cases increased force or some wiggling is required to insert the connector fully. A potential risk for patients exist, if (a) such a connector (b)
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- A Leonhard Lang letter dated June 9, 2009 was sent June 6, 2009 to its customer. The letter identified the problem and the product. The letter instructed the consignee to stop using or distributing the product and to quarantine them. It also requested them to report all quantities of product that they had in stock using the attached form.
- Quantity in commerce
- 3044
- Distribution
- Worldwide Distribution - US - FL Foreign: Austria, Germany, Romania, France, Dubai, Greece, Italy, South Korea, United Kingdom, Hungary, Latvia, Saudi Arabia and Lebanon
- Product codes and lots
- DF-20 S&W Lot #'s 80505-0778, 80814-0773, 80930-0775, 81017-0777, 90226-0776, 90403-0779, 90420-0775