Total Knee Junior Model Number 1100

Ossur Iceland ehf. / initiated 2009-08-10 / Terminated / root cause: manufacturing process

Ossur H / F began a recall of Total Knee Junior Model Number 1100 Product Usage: The Total Knee Junior is an external polycentric prosthetic knee with a geometric locking system on . The reason FDA records is "Some units of the Total Knee Junior produced between November 25th, 2008 and through April, 2009 contain balancing units that may fail under extreme use over a long period. The fault is noncritical and no incidents or...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0965-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ossur Iceland ehf.
Product
Total Knee Junior Model Number 1100 Product Usage: The Total Knee Junior is an external polycentric prosthetic knee with a geometric locking system.
Product code
ISY Joint, Knee, External Limb Component
Reason for recall
Some units of the Total Knee Junior produced between November 25th, 2008 and through April, 2009 contain balancing units that may fail under extreme use over a long period. The fault is noncritical and no incidents or injuries resulting from this situation have been reported to Ossur at this stage. Over long term use (estimated to be in the four to six year range) the prosthetic device may fail a
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Ossur sent a Safety Notice dated August 10, 2009 by mailed to all affected consignees. Ossur informed consignees of the above mentioned public reason for recall and asked consignees to take the following action: Identify the users of each of the Total Knee Junior model devices listed above and check inventory and isolate any recalled devices on hand but have not fitted on users. The letter states that Sue Fitzgerald from customer service will be contacting consignees in the next few days to coordinate replacement of the affected devices. For question call 1-(800) 219-2150, Ext: 4109. In addition, Ossur website: www.ossur.com we will be posting updates as appropriate.
Quantity in commerce
141 units
Distribution
Worldwide Distribution - USA (nationwide) in the state of: AL, AR, AZ, CA, CO, DE, FL, GA, IL, IN, KS, KY, LA, MA, MI, MN, MO, NC, NJ, NV, NY, OH, OK, OR, PA, SC, TX, and WV and the countries of: Australia, Canada, China, France, Germany, Hong Kong, Israel, Japan, Malaysia, Netherlands, New Zealand, Portugal, Puerto Rico, Saudi Arabia, South Africa, Spain, Sweden, Taiwan, United Kingdom and United States.
Product codes and lots
Serial Numbers: 8573, 8574, 8575, 8577, 8578, 8580, 8581, 8582, 8583, 8584, 8585, 8586, 8587, 8588, 8589, 8599, 8600, 8602, 8603, 8604, 8605, 8606, 8609, 8610, 8611, 8613, 8616, 8620, 8621, 8622, 8624, 8628, 8629, 8630, 8631, 8632, 8633, 8634, 8635, 8636, 8638, 8639, 8640, 8642, 8643, 8644, 8645, 8646, 8666, 8670, 8671, 8672, 8673, 8674, 8675, 8678, 8679, 8680, 8682, 8684, 8685, 8687, 8688, 8691, 8692, 8693, 8694, 8695, 8696, 8697, 8698, 8699, 8701, 8715, 8716, 8717, 8719, 8720, 8722, 8723, 8725, 8739, 8755, 8756, 8757, 8758, 8759, 8760, 8761, 8784, 8787.