Z-2075-2009

Medtronic Intrathecal Catheter Pump Segment Revision Kit, model 8596SC, Method of Sterilization: Ethylene...

Medtronic Neuromodulation / initiated 2009-08-27 / Terminated / root cause: design

Medtronic Neuromodulation began a recall of Medtronic Intrathecal Catheter Pump Segment Revision Kit, model 8596SC, Method of Sterilization: Ethylene Oxide. Medtronic, Inc., Minneapolis, MN 55432-5604... on . The reason FDA records is "Current labeling for Medtronic Sutureless Connector (SC) catheters and revision kits (hereafter referred to as SC catheters) incorrectly states that SC catheters are compatible with Medtronic IsoMed constant-flow...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2075-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Neuromodulation
Product
Medtronic Intrathecal Catheter Pump Segment Revision Kit, model 8596SC, Method of Sterilization: Ethylene Oxide. Medtronic, Inc., Minneapolis, MN 55432-5604 USA. Contents: 66-cm-pump segment with attached sutureless pump connector, Spinal segment strain-relief sleeves, Pump Segment strain-relief sleeves, connector pin. This is a component of an implantable drug infusion system, which consists of a drug infusion pump and a catheter. The drug infusion system stores and delivers parenteral drugs to the intrathecal space.
Product code
LKK Pump, Infusion, Implanted, Programmable
Reason for recall
Current labeling for Medtronic Sutureless Connector (SC) catheters and revision kits (hereafter referred to as SC catheters) incorrectly states that SC catheters are compatible with Medtronic IsoMed constant-flow infusion pumps. SC catheters are compatible with Medtronic SynchroMed II and SynchroMed EL pumps. Medtronic has determined that SC catheters are not compatible with IsoMed pumps.
Root cause tag
design
FDA root cause
Device Design
Action
A "Medical Device Correction Notification" letter was issued on August 14, 2009 to implanting and managing physicians. The letter described the nature of the issue, scope of affected product, potential severity of the issue and provided recommendations. Physicians were instructed that Sutureless Connector (SC) catheters must not be used with IsoMed pumps for future revisions and implants. A copy of the letter was also sent to the Risk Manager at each hospital associated with physicians using SC catheters. Healthcare professionals were asked to complete and return the enclosed reply card. Direct questions to your Medtronic field representative or contact Medtronic Neuromodulation Technical Services at 1-800-707-0933. Patient Management information is also available at http://www.professional.medtronic.com under the Heading, "Advisories".
Quantity in commerce
20,000 of all products
Distribution
Worldwide Distribution -- US (including Washington D.C. and Puerto Rico), Algeria, Aruba, Australia, Austria, Belgium, Canada, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Luxembourg, Malta, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Norway, Poland, Portugal, Reunion, Romania, San Marino, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom and Vatican City.
Product codes and lots
Model 8596SC used with all IsoMed Pump Models including 8472. Not lot or serial number specific.