Z-0168-2010

Guardian Services, Part Number 0240-009-430 with Part Numbers: 0240-030-004 and/or 0240-030-002, 0240-030-008...

Stryker Imaging / initiated 2009-07-31 / Terminated

Stryker Imaging began a recall of Guardian Services, Part Number 0240-009-430 with Part Numbers: 0240-030-004 and/or 0240-030-002, 0240-030-008 (Net Practice). The intended us of Guardian is to... on . The reason FDA records is "Product serving as data backup protection does not contain 100% data backup for all NetPractice customers on a daily basis". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0168-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Imaging
Product
Guardian Services, Part Number 0240-009-430 with Part Numbers: 0240-030-004 and/or 0240-030-002, 0240-030-008 (Net Practice). The intended us of Guardian is to act as a medical data backup and disaster recovery plan.
Product code
LMB Device, Digital Image Storage, Radiological
Reason for recall
Product serving as data backup protection does not contain 100% data backup for all NetPractice customers on a daily basis.
FDA root cause
Other
Action
Firm notified consignees by an Urgent: Device Correction letter, dated July 31, 2009. The letter states the problem and that Stryker is upgrading its internal infrastructure to remediate the problem. The letter also provides Stryker's recommendations for reducing risk to patients. Technical instructions on the backup storage for the NetPractice are included with the letter. Stryker will not bill customers for the Guardian Service until the problem is remediated and Stryker notifies customers. If there are questions, customers should contact Stryker at 972-410-7100.
Quantity in commerce
55 devices affected
Distribution
Nationwide Distribution
Product codes and lots
All accounts that have NetPractice and are signed up for Guardian Services backup.