Z-0160-2010
Guardian Services, Guardian Software, Part Number 0240-009-430 (Guardian Services) with Part Numbers...
Stryker Imaging / initiated 2009-08-03 / Terminated
Stryker Imaging began a recall of Guardian Services, Guardian Software, Part Number 0240-009-430 (Guardian Services) with Part Numbers 0250-003-016, 0250-003-017 and 0250-003-018 (OrthoPad)... on . The reason FDA records is "Guardian Services data backup does not contain 100% data for all OrthoPad customers on a daily basis". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0160-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Imaging
- Product
- Guardian Services, Guardian Software, Part Number 0240-009-430 (Guardian Services) with Part Numbers 0250-003-016, 0250-003-017 and 0250-003-018 (OrthoPad) Intended use of Guardian Services is to act as a medical data backup and disaster recovery plan.
- Product code
- LMB Device, Digital Image Storage, Radiological
- Reason for recall
- Guardian Services data backup does not contain 100% data for all OrthoPad customers on a daily basis.
- FDA root cause
- Other
- Action
- Firm notified consignees by an Urgent: Device Correction letter, dated 7/31/2009. The letter states the problem and that Stryker is upgrading its internal infrastructure to remediate the problem. The letter also states the firm's recommendations to reduce risk to patients. Stryker will not bill customers for the Guardian service until the problem is remediated and the firm notifies customers. If there are any questions, customers are to contact the firm at 972-410-7100.
- Quantity in commerce
- 85
- Distribution
- Nationwide Distribution
- Product codes and lots
- All accounts that have OrthoPad and are signed up for Guardian Services backup.