Z-1708-2009

Sterile, Drape, Headset, Upper, Part Number 888945-01

GE Healthcare / initiated 2008-08-15 / Terminated / root cause: packaging

GE Medical Systems, LLC began a recall of Sterile, Drape, Headset, Upper, Part Number 888945-01 on . The reason FDA records is "GE Healthcare has found that the sterile products manufactured by Contour Fabricators, Inc. (CFI) for GE healthcare (GEHC) may have seals that could be compromised during the shelf life of the products. This issue could...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1708-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
Sterile, Drape, Headset, Upper, Part Number 888945-01.
Product code
MMP Cover, Barrier, Protective
Reason for recall
GE Healthcare has found that the sterile products manufactured by Contour Fabricators, Inc. (CFI) for GE healthcare (GEHC) may have seals that could be compromised during the shelf life of the products. This issue could result in the product not maintaining its sterile state. Eleven complaints and 3 MDR's have been filed. No illness or injuries reported.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
GE Healthcare issued an Urgent Recall Notice dated August 15, 2008. The letter described the problem, provided a list of products, instructed customers to stop using the products and to return or dispose. To receive credit, customers must complete the attached product return form and include it in the package with returned items. All customers were asked to complete and fax back the attached response form.
Quantity in commerce
50 each
Distribution
Nationwide Distribution -- AZ, CA, FL, MD, NY, OR, PA, SC and TX.
Product codes and lots
all lots