Z-0134-2010
Stryker Imaging, Guardian Services Software, Part Number 0240-009-430 with Part Numbers 0240-020-013 and...
Stryker Imaging / initiated 2009-08-31 / Terminated
Stryker Imaging began a recall of Stryker Imaging, Guardian Services Software, Part Number 0240-009-430 with Part Numbers 0240-020-013 and 0240-020-022 (Office PACS Power). Intended use is to... on . The reason FDA records is "Firm has determined Guardian Services does not contain 100% data backup for 12 OfficePACS Power customers on a daily basis". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0134-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Imaging
- Product
- Stryker Imaging, Guardian Services Software, Part Number 0240-009-430 with Part Numbers 0240-020-013 and 0240-020-022 (Office PACS Power). Intended use is to act as a medical data backup and disaster recovery plan.
- Product code
- LMB Device, Digital Image Storage, Radiological
- Reason for recall
- Firm has determined Guardian Services does not contain 100% data backup for 12 OfficePACS Power customers on a daily basis.
- FDA root cause
- Other
- Action
- Stryker Imaging issued an "Urgent: Device Correction" notice dated August 18, 2009. Customers were notified of the defective product and informed that Stryker will be in contact shortly to coordinate the product correction. Actions to be taken during the interim were also detailed. For further information, contact Stryker Imaging at 1-972-410-7100.
- Quantity in commerce
- 12 devices affected
- Distribution
- Nationwide Distribution -- AR, TN, IL, GA, TX, WA, MI, PA, and MS.
- Product codes and lots
- All accounts that have OfficePACS Power and are enrolled in Guardian Services backup.