Stryker Infravision Esophageal Kit, Model Numbers 220180548, 220180540, 220180560, manufactured by Stryker...
Stryker Corp. / initiated 2009-08-20 / Terminated / root cause: design
Stryker Endoscopy began a recall of Stryker Infravision Esophageal Kit, Model Numbers 220180548, 220180540, 220180560, manufactured by Stryker Endoscopy San Jose on . The reason FDA records is "The silicone sheath component of the product detaches from the catheter during removal of the e-kit from the body". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0830-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Corp.
- Product
- Stryker Infravision Esophageal Kit, Model Numbers 220180548, 220180540, 220180560, manufactured by Stryker Endoscopy San Jose.
- Product code
- FAT Bougie, Esophageal, And Gastrointestinal, Gastro-Urology
- Reason for recall
- The silicone sheath component of the product detaches from the catheter during removal of the e-kit from the body.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Stryker issued Urgent Product Recall letters dated August 20, 2009 to all its direct consignees, informing them of the affected products and providing instructions on the recall. Customers were requested to return the affected product to Styrker.
- Quantity in commerce
- 2985 units
- Distribution
- Worldwide Distribution: USA in the states of: Alabama, Arkansas, California, Colorado, Connecticut, Delaware, Florida, Georgia, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Michigan, Minnesota, Missouri, North Carolina, New Hampshire, Nebraska, New Jersey, New Mexico, New York, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Virginia, Vermont, Washington, Wisconsin, West Virginia, and Wyoming, and the countries/region of Australia, Canada, Scandinavia, Italy, and Switzerland.
- Product codes and lots
- 5687,5766,5815,5822,5833,5872, 5924,60973888,60987814, 61004943, STROll409-l, STROl1409-2, STRO 11609-1, STROl1609-2, STR02l009-1, STR02l009-8, STR022709-l, STR022709-2, STR03ll09-1, STR03ll09-3, STR040l09-l, STR041709-2, STR041709-4, STR0430009-2, STRlO0208-l, STR102208-1, STR102408-l, STRI02708-7, STR103l08-1, STRlll308-l, STR12l808-l, STR12l808-2, STROl1209-l