Vacuum Collection Bottle, 1/2 gallon (1300 mL); Ohio Medical Company, Gurnee, IL 60031; part number...
Ohio Medical Corporation began a recall of Vacuum Collection Bottle, 1/2 gallon (1300 mL); Ohio Medical Company, Gurnee, IL 60031; part number 0212-0301-600 (6700-0206-500); the bottles are sold alone... on . The reason FDA records is "The re-useable collection bottles were made with polycarbonate instead of the specified polysulfone. The recommended cleaning of the bottles include use of ammonia and autoclaving, which may have an adverse impact on...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0089-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ohio Medical LLC
- Product
- Vacuum Collection Bottle, 1/2 gallon (1300 mL); Ohio Medical Company, Gurnee, IL 60031; part number 0212-0301-600 (6700-0206-500); the bottles are sold alone and with various cap & float assemblies including: 0321-7986-801 - Bottle with Straight Wall Bracket, Cap & Float; 6700-0019-800 - Bottle with Floor Bracket, Cap & Float; 6700-0024-800 - Bottle with V Wall Bracket, Cap & Float. The collection bottles are part of a suction bottle assembly which is designed to collect various body fluids.
- Product code
- KDQ Bottle, Collection, Vacuum
- Reason for recall
- The re-useable collection bottles were made with polycarbonate instead of the specified polysulfone. The recommended cleaning of the bottles include use of ammonia and autoclaving, which may have an adverse impact on polycarbonate.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Ohio Medical Corporation issued recall letters beginning September 17, 2009. Consignees were informed of the affected product and requested to return affected collection bottles to the firm. The accounts were requested to complete and return the enclosed acknowledgement form, indicating the number of affected units on hand, to obtain a returned material authorization for the return of the recalled bottles and their replacement. For further information, contact Ohio Medical Quality Assurance Manager, Anita Christ, at 847-855-6270.
- Quantity in commerce
- 5,892 bottles
- Distribution
- Worldwide Distribution -- United States, Canada, Chile, Costa Rica, Cyprus, Egypt, El Salvador, Greece, Hong Kong, India, Israel, Japan, Jordan, Korea, Kuwait, Malaysia, Netherlands, Oman, Paraguay, Philippines, Qatar, Saudi Arabia, Singapore, Thailand, Turkey, United Kingdom and Venezuela.
- Product codes and lots
- The bottles have a date stamp on the bottom which consists of a clock dial and a two digit number in the middle with an arrow pointing to the hour number on the dial. The codes 07 -> 5 and 08 ->10 are affected.
FIRST ARTICLE / WEEKLY