Z-0090-2010

Vacuum Collection Bottle, 1 gallon (3200 mL); Ohio Medical Company, Gurnee, IL 60031; part number...

Ohio Medical LLC / initiated 2009-09-21 / Terminated / root cause: manufacturing process

Ohio Medical Corporation began a recall of Vacuum Collection Bottle, 1 gallon (3200 mL); Ohio Medical Company, Gurnee, IL 60031; part number 0212-0302-600 (6700-209-500); the bottles are sold alone and... on . The reason FDA records is "The re-useable collection bottles were made with polycarbonate instead of the specified polysulfone. The recommended cleaning of the bottles include use of ammonia and autoclaving, which may have an adverse impact on...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0090-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ohio Medical LLC
Product
Vacuum Collection Bottle, 1 gallon (3200 mL); Ohio Medical Company, Gurnee, IL 60031; part number 0212-0302-600 (6700-209-500); the bottles are sold alone and with various cap & float assemblies: 0321-7986-800 - Bottle with Straight Wall Bracket, Cap & Float; 6700-0018-800 - Bottle with V Wall Bracket, Cap & Float; 6700-0370-800 - Bottle with Floor Bracket, Cap & Float. The collection bottles are part of a suction bottle assembly which is designed to collect various body fluids.
Product code
KDQ Bottle, Collection, Vacuum
Reason for recall
The re-useable collection bottles were made with polycarbonate instead of the specified polysulfone. The recommended cleaning of the bottles include use of ammonia and autoclaving, which may have an adverse impact on polycarbonate.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Ohio Medical Corporation issued recall letters beginning September 17, 2009. Consignees were informed of the affected product and requested to return affected collection bottles to the firm. The accounts were requested to complete and return the enclosed acknowledgement form, indicating the number of affected units on hand, to obtain a returned material authorization for the return of the recalled bottles and their replacement. For further information, contact Ohio Medical Quality Assurance Manager, Anita Christ, at 847-855-6270.
Quantity in commerce
823 bottles
Distribution
Worldwide Distribution -- United States, Canada, Chile, Costa Rica, Cyprus, Egypt, El Salvador, Greece, Hong Kong, India, Israel, Japan, Jordan, Korea, Kuwait, Malaysia, Netherlands, Oman, Paraguay, Philippines, Qatar, Saudi Arabia, Singapore, Thailand, Turkey, United Kingdom and Venezuela.
Product codes and lots
The bottles are coded with a date stamp on the bottom of the bottle. The affected code is an "E" inside of a circle.