Z-0654-2010

Complete SE Biliary Stent System, SC5120L (5 mm x 12 mm), lot number 0000983799, UPC code 00613994186478, use...

Medtronic, Inc. / initiated 2009-07-17 / Terminated / root cause: labeling and UDI

Medtronic CardioVascular began a recall of Complete SE Biliary Stent System, SC5120L (5 mm x 12 mm), lot number 0000983799, UPC code 00613994186478, use by date 03/17/2011, manufactured by Medtronic... on . The reason FDA records is "Mislabeling-- A lot of 5 mm x 120 mm stents were labeled as 6 mm x 80 mm stents and vice versa". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0654-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Complete SE Biliary Stent System, SC5120L (5 mm x 12 mm), lot number 0000983799, UPC code 00613994186478, use by date 03/17/2011, manufactured by Medtronic CardioVascular, Santa Rosa, CA. Indicated for use in the palliation of malignant neoplasms in the biliary tree.
Product code
FGE Stents, Drains And Dilators For The Biliary Ducts
Reason for recall
Mislabeling-- A lot of 5 mm x 120 mm stents were labeled as 6 mm x 80 mm stents and vice versa.
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
Medtronic, Inc. issued an "Urgent - Voluntary Medical Device Recall" communication dated July 17, 2009 asking consignees to segregate labeled lot numbers until a representative from the firm arrived to retrieve the products. For further information, contact your local Medtronic representative or call Medtronic, Inc. at 1-707-525-0111.
Quantity in commerce
20 units
Distribution
Product was distributed to 24 consignees in TX, MO, PA, MN, NY, IL, CA, IN, AZ, TN, OH, MI, and CT.
Product codes and lots
Lot Number: 0000983799, use by date 03/17/2011.